Voltage |
8 ranges: 0.3 V, 1 V, 3 V, 10 V,
30 V, 100 V, 300 V, 1000 V |
|
Frequency range |
DC, 0.1 Hz-1 MHz |
Crest Factor |
3:1 at 50 % full scale (fs) |
Input Impedance |
1 MOhm |
Common Mode 50 Hz/100 kHz |
160 dB/100 dB |
Standard accuracy 23°C;
rms, mean, rectified mean; 0.3V typical
1 Hz-1 kHz ±(0.1 % rdg +0.1 % range)
DC, 1 kHz-10 kHz ±(0.2 % rdg +0.2 % range)
10 kHz-100 kHz
±(0.3 %1 range + 0.04 % /kHz rdg)
100 kHz-300 kHz
±(0.3 %/ range + 0.04 % /kHz rdg), typical |
Improved accuracy
±(0.05 % rdg + 0.07 % range) |
Current |
13 ranges: 1.5 mA, 5 mA, 15 mA, 50 mA, 150 mA, 500 mA. 1.5 A, 5 A; 1, 3, 10, 30, 100 A. |
Max. 1 A, 5 A, 30 A, resp. |
Frequency range |
DC, 0.1 Hz-300 kHz / 1 MHz |
Crest Factor |
3:1 at 50 % full scale (fs) |
Common Mode 50 Hz/100 kHz |
160 dB/120 dB |
Standard accuracy 23°C; 1 A-, 5 A-, shunt input 30 A input
1 Hz-1 kHz ±(0.1%rdg +0.1 % rng)
±(0.1%rdg+0.1 % rng)
DC, 1 kHz-10 kHz ±(0.2 % rdg + 0.2 % rng)
±(0.7 % rdg + 0.2 % rng)
10 kHz-100 kHz #0.3 % range+ 0.04 %/kHz rdg) ±(0.3 % rng + 0.5 %/kHz rdg), typical
100 kHz-300 kHz ±(0.3 % range + 0.04 %/kHz rdg), typical |
Lowest ranges 1.5 mA, 15 mA,
1 A: typical.
Improved accuracy
1 Hz-400 Hz
±(0.05 % rdg + 0.07 % range) |
Power |
104 ranges corresponding to the products V x A. |
|
Frequency range |
DC, 0.1 Hz-300 kHz |
45 Hz-65 Hz (0.1 % rdg + 0.01 % range)
1 Hz-1 kHz Add accuracy percentage figures of current and voltage,
DC, 1 kHz-10 kHz +0.04 %/kHz PF
10 kHz-100 kHz |
PF= 0 to ±0.1
PF= 0 to ±1
PF= 0 to ±1
PF=1 |