Force Capacity |
± 10 N, 1 kgf, 2.25 lbf |
Mechanical Fitting |
M10 × 1.5RH central thread |
Linearity |
±0.25% of Reading from 0.4 to 100% of Force Capacity |
Repeatability |
0.25% of Reading from 0.4 to 100% of Force Capacity |
Hysteresis |
±0.1% of Maximum Reading in cycle |
Creep |
±0.1% of Reading (5 Seconds to 3 Minutes) |
Max Tare Weight |
40% of Force Capacity |
Overload |
150% of Force Capacity without calibration change
300% of Force Capacity without mechanical failure |
Off-Center Error |
±0.5% of Reading (10 mm offset) |
Frame Compatibility |
5900 and 3300 |