HP Agilent 54624A Digitizing Oscilloscope 100 MHz, 200 MSa/s

Rent HP Agilent 54624A Digitizing Oscilloscope 100 MHz, 200 MSa/s
Enlarge Image
Request Quote Print Page
Manual Datasheet

The HP Agilent 54624A Digitizing Oscilloscope 100 MHz, 200 MSa/s will give you the channel count and measurement power you need, including MegaZoom deep memory, high definition display and flexible triggering, especially if your designs include heavy analog content. Whether youre testing designs with four inputs, such as anti-lock brakes, or monitoring multiple outputs of a power supply, the four-channel model helps you get your debug and verification done with ease.

 

Features:

  • 100-MHz Bandwidth Oscilloscope
  • 4 channels
  • Maximum Memory 4 MB
  • Maximum Sample Rate: 200 MSa/s

 

Specifications:

Acquisition: Analog Channels
Max Sample Rate 200 MSa/s
Max Memory Depth 4 MB interleaved,
Vertical Resolution 8 bits
Peak Detection 5 ns
Averaging Selectable from 2, 4, 8, 16, 32, 64 … to 16383
High Resolution Mode 12 bits of resolution when ≥500 μs/div, (average mode with ave = 1)
Filter Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever is less) with vectors on
Vertical System: Analog Channels
Analog Channels Ch 1, 2, 3 and 4 simultaneous acquisition
Bandwidth (-3dB) dc to 500 MHz
ac Coupled 3.5 Hz to 500 MHz
Calculated Risetime
=0.35/bandwidth)
~700 ps
Single Shot Bandwidth 500 MHz maximum
Range1,2 1 mV/div to 5 V/div
Maximum Input CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50 Ω input
Offset Range ±5 V on ranges <10 mV/div; ±25 V on ranges 10 mV/div to 199 mV/div;±100 V on ranges ≥200 mV/div
Dynamic Range Lesser of ±8 div or ±32 V from center screen
Input Resistance 1 MΩ ±1%, 50 Ω selectable
Input Capacitance ~ 14 pF
Coupling ac, dc, ground
>BW Limit ~ 20 MHz selectable
Channel-to-Channel Isolation
(with channels at same V/div)
dc to 20 MHz > 40 dB; 20 MHz to max bandwidth > 30 dB
ESD Tolerance ±2 kV
Noise Peak-to-Peak 2% full scale or 1 mV, whichever is greater
Common Mode Rejection Ratio 20 dB @ 50 MHz
DC Vertical Gain Accuracy*1,2 ±2.0% full scale
DC Vertical Offset Accuracy <200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset value; ≥200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value
Single Cursor Accuracy1,2 ±{DC Vertical Gain Accuracy + DC Vertical Offset Accuracy + 0.2% full scale (~1/2 LSB) }
Dual Cursor Accuracy*1,2 ±{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)}
Horizontal
Range 5 ns/div to 50 s/div
Resolution 25 ps
   
Vernier 1-2-5 increments when off, ~25 minor increments between major settings when on
Delay Range Pre-trigger (negative delay): Greater of 1 screen width or 10 ms
Post-trigger (positive delay): 500 seconds
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

1 For 54621A, 54621D, 54622A, 54622D, and 54624A, 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting. 2 For 54641A, 54641D, 54642A, and 54642D, 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.