Keysight 8712B 1.3 GHz RF Vector Network Analyzer

HP Agilent 8712B 1.3 GHz RF Vector Network Analyzer
Enlarge Image
HP Agilent 8712B 1.3 GHz RF Vector Network Analyzer HP Agilent 8712B 1.3 GHz RF Vector Network Analyzer
Keysight 8712B
Request Quote Print Page
Manual Datasheet
300 kHz - 1300 MHz, RF Vector Network Analyzer
 
Options
  • 1EC - 75 ohms
  • 1DA/1DB - AM delay
  • 1C2 - IBASIC
  • 1E1 - Step attenuator
  • 100 - Fault location and structural return loss software
  • K02 (Special Option) - Switching test sets
 
Keysight 8712B Specs
Source
Frequency
  • Range: 300 kHz to 1.3 GHz
  • Resolution: 1 Hz
  • Stability: ±5 ppm 0° C to 55° C typical
  • Accuracy:
    • 1) ±5 ppm @ 25° C ±5° C
    • 2) <1 Hz at 10% change in line voltage
  • Harmonics:
    • <-20 dBc, <1MHz
    • <-30 dBc, >1MHz
  • Output Power:
    • Resolution: 0.01 dB
    • Level Accuracy: ±1.0 dB
 
Receiver
Frequency Range
  • Narrowband: 300 kHz to 1.3 GHz
  • Broadband: 0.10 to 1.3 GHz

Dynamic Range

  • Narrowband (50 ohms):
    • >100 dB, ≥5 MHz (+10 to -90 dBm)
    • >60 dB <5 MHz (+10 to -50 dBm)
  • Narrowband (75 ohms):
    • >97 dB, >5 MHz (+10 to -87 dBm)
    • >57 dB, <5 MHz (+10 to -50 dBm)
  • Broadband (50 ohms): >66 dB (+16 to -50 dBm)
  • 75 ohms: >63 dB (+16 to -47 dBm)

Damage Level: -23 dBm, ±25 VDC

Maximum Input

  • Narrowband (0.5 dB compression): +10 dBm
  • Broadband (0.55 dB compression): +16 dBm
 
Measurements Ports
  • Directivity: 40 dB
  • Source Match (reflective): 20 dB
  • Source Match (transmission): 14 dB typical
  • Load Match: 18 dB typical
  • Reflective Tracking: 0 ±0.4 dB typical
 
Maximum and Minimum Power
>Max.
  ≤1.0 GHz >1.0 GHz
  Power (dBm) Power (dBm)
Options Min. Min. Max.
None 0 16 0 13
1E1 -60 15 -60 12
1EC -3 13 -3 10
1DA/DB -2 14 -2 11
1E1 & 1EC -60 12 -60 9
1E1 & 1DA -60 13 -60 10
1EC & 1DB -5 11 -5 8
1EC, 1E1 & 1DB -60 10 -60 7