Keytek EMC Pro Advanced EMC Test System

Keytek EMC Pro Advanced EMC Test System
Enlarge Image
Request Quote Print Page
Specifications

The answer to manufacturers' demand for a mid-range, multi-capability EMC immunity tester, EMCPro® is easily configured to meet today's immunity standards required for CE Marking and compliance requirements. Operating via Thermo KeyTek's easy to use Windows®-based PC software or from the front panel, it enables test engineers and technicians to run compliance-level EMC immunity tests in accordance with European Norm requirements. Portable and low cost, the EMCPro® is ideal for companies who require flexibility, versatility, and the highest test level-to-cost ratio instrument on the market.
 

Configurations for the Keytek EMCPro:

Model PRO-SURGE
Surge per IEC 61000-4-5. Will do part of UL1449
Voltage Waveform 1.2/50µs
Peak Voltage 250V - 6.6kV ±5%, 12Ω Mode
250V - 6.0kV ±5%, 2Ω mode
Peak Current 125 A - 3.3 kA ±10%
Additional 10 ohm Resistor Software selectable
Repetition Rate Up to 4 per minute
Open-circuit Voltage Front time: 1.2µs ±30%
Duration: 50µs ±20%
Undershoot: ≤30%
Line sync accuracy ±15%, 50 - 277 VAC
Model PRO-EFT
EFT per IEC 61000-4-4 and ANSI
Voltage Waveform 5/50 ns ±30%
Peak Voltage 250V - 4.4 kV ±5%
Burst Period 300 ms ±10%
Burst Duration 15 ms ±20%
Frequency 1 kHz - 100 kHz in 0.5 kHz steps, ±10%
DC Blocking Capacitor 10 nF (internal)
Model PRO-RING*
Surge Ring Wave per ANSI and UL. Will do part of 61000-4-12.
Voltage Waveform 100 kHz damped cosine
Peak Voltage 250 - 6.6kV ±5%
Repetition Rate <4/minute at 6 kV, faster at lower voltages
Open-circuit Voltage Rise Time: 0.5µs ±30%
Short-circuit Current Vp/Ip: 12 ohm ±3 ohm or 30 ohm ±8 ohm software selectable
*Pro-Telecom and Pro-Ring can not be installed together in the same unit
Surge Waveform Monitoring
Lines Monitored Monitors are automatically switched to match the generator coupling mode
Open-circuit Voltage 1000:1 ±10%
Short-circuit Current Attenuation: 200:1 ±7%
Model PRO-HPOWER
Power Frequency Magnetic Field per 61000-4-8
Field Frequency 50Hz/60Hz
Field Amplitude 0.5-4A/m, in 0.25A steps, ±10% (with CM-HCOIL) up to 100A/m with the optional external HPOWER-EXT
AC Source Internal
Resolution 0.25 Amp minimum
Coil Factor 0.65 to 1.00
Coil Resistance 0.05Ω maximum
Model PRO-HPULSE
Pulse Magnetic Field per IEC 61000-4-9
Field Pulse 8/20µs
Field Amplitude 100A/m -1000A/m, ±10%
Resolution 5A/m
Coil Factor 0.65 to 1.00
Model PRO-PQF
Dips and Interrupts per IEC 61000-4-11
Dips 70%, 40%
Interrupts 0% (short and open)
Transition Time 1µs - 5µs
Inrush Minimum: 250 Amps @ 100V - 120V
Minimum: 50 0Amps @ 220V - 240V
AC Voltage 50 VAC - 250 VAC, 50 Hz - 60 Hz
AC Current 16A Max
PQF Sync Output 5V signal occurs at each Dip or interrupt transition
Dip Duration 10 msec min at 60 Hz
2 msec min at 50 Hz
PQF Waveform Monitoring
Voltage Input Connection Fixed, L1 to L2
Voltage Attenuation 100:1 ±5%
Current Input Connection Fixed, L1
Peak Current Minimum 500A inrush into 1700µF
Current Attenuation 200:1 ±5%
Model PRO-BASE
System Voltage 90 VAC - 240 VAC, 50 Hz - 60 Hz
Integrated EUT Mains Coupler/Decoupler AC Voltage: 1 phase, 50-250VAC, 50/60Hz
AC Current: 16A max
DC Voltage: 100 VDC max
DC Current: 10A max
Frequency: 50/60Hz
EUT Connectors: Nema, British, Schuko
Control Interface: RS232 Fiber-optic
Safety Features: External Interlock for users
Interlock for CCL connector
External stop input