Tektronix TDS794D Digital Phosphor Oscilloscope 2 GHz, 1 GS/s

Rent Tektronix TDS794D Digital Phosphor Oscilloscope 2 GHz, 1 GS/s
Enlarge Image
Request Quote Print Page
Datasheet

Specifications:
 

  • Bandwidth: 2 GHz
  • Number of Channels: 4
  • Number of Samplers: 4
  • Max Real-time Sample Rate:
    1 Channel: 4 GS/s
    2 Channels: 2 GS/s
    3-4 Channels: 1 GS/s
    Equivalent-time Sample Rate: 250 GS/s max
  • Max Record Length:
    1 Channel: 50 K (Opt. 1M: 500 K, Opt. 2M: 8 M)
    2 Channels: 50 K (Opt. 1M: 250 K, Opt. 2M: 4 M)
    3-4 Channels: 50 K (Opt. 1M: 130 K, Opt. 2M: 2 M)
  • Max. Sample Rate Window: 2 ms
  • Sensitivity: 10 mV/div to 1 V/div (50 ohm mode)
  • DC Gain Accuracy: ±1.0% (0.7% typical)
  • Effective Bits (typical): 5.0 (2 GHz @ 4 GS/s), 9.7 with Hi-res (1 MHz @ 10 MS/s)
  • Vertical Resolution: 8-Bits (256 levels on 10.25 divisions), >11-Bits with averaging, >13-Bits typical with Hi-res
  • Position Range: ±5 divisions
  • Offset Range: ±1 V from 1 mV to 100 mV/div, ±10 V from 101 mV to 1 V/div, ±0.5 V from 10 mV to 50 mV/div, ±0.25 V from 50.5 mV to 99.5 mV/div, ±5 V from 100 mV to 500 mV/div, ±2.5 V from 505 mV to 1 V/div
  • Analog Bandwidth Selections: Full only
  • Input Coupling: DC, GND
  • Input Impedance Selections: 1 M ohm in parallel with 10 pF, or 50 ohm only (DC coupling only)
  • Channel Isolation: >100:1 at 100 MHz and >30:1 at the rated bandwidth
  • Max Input Voltage: 5 Vrms with peaks ± 20 Volts


Applications:
 

  • Communication Compliance Testing
  • Digital Design and Debugging
  • Jitter and Timing Analysis
  • Fast Edge Characterization
  • Video Design and Debug
  • Disk Drive Measurements
  • Power Measurements

Hardcopy Outputs:
  • Printer, Plotter (HPGL), Data (MathCad, spreadsheets format), Interface (GPIB standard), Hardcopy Interface (Centronics and RS-232 (talk only))


Storage:
 

  • Non-volatile Waveform Storage: 4 full 50 K records (Opt. 1M or 2M: 4 full 130 K records, 2 full 250 K records, or 1 compressed 500 K record)
  • Non-volatile Storage for Setups: 10 front panel setups
  • Floppy Disk Drive: Store reference waveforms, setups, and image files on 3.5 in 1.44 MB or 720 K MS DOS-formatted floppy disks
  • Iomega Zip and Zip Plus Drive Compatible


Probes are optional.
 

  • Please contact us with your probe requirement.