Ranges |
Min: 0 - 50 ppb full scale
Max: 0 - 20,000 ppb full scale (selectable, dual-range supported) |
Measurement Units |
ppb, ppm, μg/m3, mg/m3 (selectable) |
Zero Noise |
< 0.2 ppb (RMS) |
Span Noise |
< 0.5% of reading (RMS) above 50 ppb |
Lower Detectable Limit |
< 0.4 ppb |
Zero Drift |
< 0.5 ppb/24 hours |
Span Drift |
< 0.5% of full scale/24 hours |
Lag Time |
20 seconds |
Rise/Fall Time |
< 100 seconds to 95% |
Linearity |
1% of full scale |
Precision |
0.5% of reading above 50 ppb |
Sample Flow Rate |
650 cc/min ±10% |
Power Requirements |
100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges |
10V, 5V, 1V, 0.1V (selectable) |
Recorder Offset |
±10% |
Included I/O |
1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device ports
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs |
Optional I/O |
1 x USB com port
1 x RS485
8 x analog inputs (0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
3 x 4 - 20mA current outputs |
Operating Temperature Range |
5 - 40ºC (with US EPA Approval) |
Dimensions (HxWxD) |
7” x 17” x 23.5” (178 x 432 x 597 mm) |
Weight |
35.7 lbs (16.2 kg) |
Certifications
All specifications are based on constant conditions. |
US EPA: EQSA-0495-100
EU: EN14212 TÜV Rheinland
QAL1 Certified: EN15267
MCerts: Sira MC0500067/07
CNEMC: 质(认)字 No. 2015-028 Report |