Teseq DCP 0100A Dual Directional Coupler


Need Help? Call (800) 404-2832ATEC AssuranceWorldwide Delivery
Teseq DCP 0100A Dual Directional Coupler
Enlarge Image
Teseq DCP 0100A
DCP 0100A Datasheet
The Teseq DCP 0100A Dual Directional Coupler is a coupler designed to augment immunity test systems by enhancing the monitoring of forward and reflected power into antennas, clamps or other transducer devices. Dual-directional and high-powered, the Teseq DCP 0100A is designed specifically for the testing of radiated immunity systems and BCI (Brain Computer Interfaces).
 
 
Teseq DCP 0100A Features
  • 4 kHz to 1 GHz Frequency Range
  • 100 W Nominal Power
  • 50 Ω Impedance
Teseq DCP 0100A Specs
Frequency Range 4 kHz to 1 GHz
Impedance 50 Ω
Power, Nominal 100 W
Insertion Loss (gate 1 - gate 2): 0.8 dB (≤ 0.5 dB up to 400 MHz/ ≤ 0.8 dB up to 1 GHz)
VSWR: < 1:1.4 (gate 1 - gate 2)
Coupling Factor 40 dB ± 1.5 dB (gate 1 - gate 3 and gate 2 - gate 4)
Directivity 10 kHz to 400 MHz > 20 dB
Cross Talk Attenuation > 40 dB (gate 3 - gate 4)
Connectors N Type Female (gate 1, 3 and 4), N Type Male (gate 2)
Dimensions L x W x D 190 mm x 105 mm x 45 mm
Weight Approx. 700 g
Teseq DCP 0100A Standards
  • IEC/EN 61000-4-3
  • IEC/EN 61000-4-6
  • IEC/EN 61000-4-20
  • CISPR 24
  • EN 61000-6-1
  • EN 61000-6-2
  • IEC 60533
  • IEC 60601-1-2
  • IEC 61326
  • MIL STD 461
  • RTCA/DO 160
  • ISO 11452-2
  • ISO 11452-3
  • ISO 11452-4
  • ISO 11452-5
  • SAE J1113-4
  • SAE J1113-24