Teseq MFO 6502 Automatic Power Line Frequency Magnetic Field

Teseq MFO 6502 Automatic Power Line Frequency Magnetic Field
Enlarge Image
Teseq (Schaffner) MFO 6502
Request Quote Print Page
Manual Datasheet
Electronic devices containing components sensitive to magnetic fields need to be tested for immunity to electromagnetic fields. Product standards define the applicability of such tests and specify the field strength. The basic standard IEC/EN 61000-4-8 describes the test methods for magnetic fields with powerline frequencies.

The MFO 6502 is an auxiliary device for the NSG 3040 series and provides fully automatic power line magnetic field tests. It is connected via system cable to the NSG 3040 and is fully embedded in the configuration management, the software control and the report mechanism.

Features:

  • Integrated function unit to NSG 3040
  • Magnetic field test with 50 or 60 Hz up to 40 A/m
  • Complies with IEC/EN 61000-4-8

Specifications:

Parameter Value
With magnetic antenna loop INA 701

Field strength, low range: 0.08 to 0.4 A/m (80 to 400 mA drive current)
High range: 0.2 to 3.6 A/m (200 mA to 4.1 A drive current)

With magnetic antenna loop INA 702

Field strength, low range: 0.8 to 4 A/m (80 to 400 mA drive current)
High range: 2 to 40 A/m (200 mA to 4.1 A drive current)

Total harmonic distortion (THD)

<8% for levels 1 to 4, <3.5% at max. output

Test frequency

Selectable 50 and 60 Hz ±3%

Magnetic field adjustment

Software driven