Teseq NSG 3040 Transient Generator

Rent Teseq NSG 3040 Multifunction Generator for Surge, EFT and PQT
Enlarge Image
Rent Teseq NSG 3040 Multifunction Generator for Surge, EFT and PQT Rent Teseq NSG 3040 Multifunction Generator for Surge, EFT and PQT
Teseq (Schaffner) NSG3040
Request Quote Print Page
NSG3040 Datasheet Accessory Kit
The Teseq NSG 3040 Transient Generator generates surge voltage to 4.4 kV and emits an EFT/Burst pulse of up to 4.8 kV/1 MHz. Teseq NSG 3040 devices simulate electromagnetc interference effects for the purpose of immunity testing in compliance with manufacturer, national and international standards. Technicians can utilize this product in assignments calling for IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-8, IEC 61000-4-9, and IEC 61000-4-11 standard-compliant devices. The "Master-Slave" architecture of the units allow individual pulse modules to be calibrated separately; calibration data and correction factors can be stored on the slave controller. Teseq NSG 3040 systems are designed to fulfill EMC test requirements for CE mark testing, including combination wave surge, Power Quality Testing (PQT) and Electrical Fast Transient (EFT) pulses. The 7.2" color TFT display featurs a touchscreen interface, a wheel with step size keys and large, start-stop-pause keys.
 
Teseq (Schaffner) NSG3040 Features
  • Modular, expandable system 
  • Surge voltage to 4.4 kV
  • EFT/Burst to 4.8 kV / 1 MHz
  • PQT to 16 A / 270 VAC & DC
  • Easy to use 7“ color touch screen
  • Parameters can be changed while test is running
  • Wide range of optional test accessories
Teseq (Schaffner) NSG3040 Specs

Surge Combo Wave Pulse Pulse

Conforms to IEC/EN 61000-4-5
 
Parameter Value
Pulse voltage (open circuit) 200 V to 4.4 kV (in 1 V steps)
Pulse current (short circuit) 100 A to 2.2 kA
Impedance 2/12 Ω
Polarity Positive / negative / alternate
Pulse repetition 10 s, up to 600 s (in 1 s steps)
Test duration 1 to 9999 pulses, continuous
Phase synchronization Asynchronous, synchronous 0 to 359º (in 1º steps)
Coupling External / internal
 

Burst (EFT) 5/50 ns

Pulse conforms to IEC/EN 61000-4-4
 
Parameter Value
Pulse amplitude 200 V to 4.8 kV (in 1 V steps) - open circuit
100 V to 2.4 kV (50 Ω matching system)
Burst frequency 100 Hz to 1000 kHz
Polarity Positive / negative / alternate
Repetition time 1 ms to 4200 s (70 min)
Burst time 1 µs to 1999 s, single pulse, continuous
Test duration 1 s to 1000 h
Phase synchronization Asynchronous, synchronous 0 to 359º (in 1º steps)
Coupling External / internal
 

PQF Dips & Drops

Conforms to IEC/EN 61000-4-11, IEC/EN 61000-4-29
 
Parameter Value
Dips & drops From EUT voltage input to 0 V, 0%
Uvar with optional variac Depending on model (VAR 650x)
Uvar with step transformer 0, 40, 70, 80% (INA 650x)
Peak inrush current capability 500 A (at 230 V)
Switching times 1 to 5 μs (100 Ω load)
Event time 20 µs to 1999 s, 1 to 99’999 cycles
Test duration 1 s to 70’000 min, 1 to 99’999 events, continuous
Repetition time 40 µs to 35 min, 1 to 99’999 cycles
Phase synchronization Asynchronous, synchronous 0 to 359º (in 1º steps)
CDN 8014 / CDN 8015 Capacitive coupling clamp for burst
CDN 163 Burst coupling network 100 A per phase (coupling all to ref ground)
CDN 3083 Surge Coupling/Decoupling Network 3 Phase, 100 Amp, 690 Volt
CDN 117 / CDN 118 Coupling networks for signal-/data lines (surge)
CAS 3025 Burst/EFT verification set
MD 200 Voltage differential probe 7 kV
MD 300 Current probe 5 kA
INA 165 Conducted stand-off
INA 166 Brackets 5 HU for rack mounting

Teseq NSG 3040-04 Option:
 
  • Multi-function generator for testing EN/IEC 61000-4-4, 61000-4-5, 61000-4-11
  • Mainframe NSG 3040
  • CWM 3450
  • FTM 3425
  • PQM 3403
  • 260V/16A single phase CDN
  • LAN (TCP/IP) interface
  • Cables
  • Traceable calibration certificate
  • Win3000 software not included
Teseq (Schaffner) NSG3040 Standards
  • IEC 61000-4-4
  • IEC 61000-4-5
  • IEC 61000-4-8
  • IEC 61000-4-9
  • IEC 61000-4-11
  • IEC/EN 61000-4-29
  • EMC Directives
  • UN ECE R10 (Automotive)
  • ANSI/IEEE C62.41
  • EN 61000-6-1
  • EN 61000-6-2
  • EN 50370
  • EN 61643
  • EN 300386-2
  • EN 300386 V1.3.2
  • EN 301489-1
  • EN 301489-7
  • EN 301489-17
  • EN 301489-24
  • IEC 60255-1
  • IEC 60601-1-2
  • IEC 61326
  • IEC 61850-3
  • ITU-T K.12
  • ITU-T K.20
  • ITU-T K.21
  • ITU-T K.28
  • ITU-T K.41
  • ITU-T K.44
  • ITU-T K.45