Test Channels |
1 – 192 per system |
Voltage Ranges |
Available up to 180 V maximum |
Voltage Accuracy |
0.02% of full scale voltage |
Voltage Resolution |
16 bit |
Current Ranges |
Single current range: 1 mA – 2000 A
Four current ranges: 150 µA, 5 mA, 150 mA, 5 A |
Current Accuracy |
0.02% of full scale current on 5 Amp multi-range channels |
Current Accuracy |
0.05% of full scale current on all other channels |
Current Resolution |
16 bit |
Time Resolution |
10 mS standard with 5 mS and 1 mS as an option |
Minimum Pulse Width |
100 µS |
Rise Time |
<500 µS standard in constant current modes
100 µS or 20 µS optional |
Data Recording Rate |
200 data points per second per system standard |
Data Recording Interval |
ΔTime (minimum 10 mS standard, 5 mS and 1 mS as an option), ΔV, ΔI, ΔAh, ΔWh, ΔT, ΔP |
Operating Modes |
Constant current, constant voltage, constant power, constant resistance, voltage scan (cyclic voltametry) |
Number of Steps per Test |
127 standard plus subroutines |