Vishay 2310 Signal Conditioning Amplifier System

Vishay 2310 Signal Conditioning Amplifier System
Enlarge Image
Vishay 2310 Signal Conditioning Amplifier System Vishay 2310 Signal Conditioning Amplifier System Vishay 2310 Signal Conditioning Amplifier System
Vishay 2310
Request Quote Print Page
Manual

Features:

  • Double-shunt (for transducer) across opposite bridge arms.
  • Single-shunt (for stress analysis) across any bridge arm, including dummy gage.
  • Circuit (2-level, dual polarity):
  • Provision for 4 dedicated leads to shunt external arms.
  • Cal circuit selected by switches on PC board.
  • Standard Factory-Installed Resistors ( + 0.1%) Simulate:
          + 200 and + 1000 microstrain at GF=2 across dummy half bridge;
          +1000 microstrain at GF=2 across dummy gage (120 and 350Ω);
          + 1 mV/V (double-shunt) for 350Ω transducer.
  • Remote-Operation Relays (Option Y):
    4 relays (plus remote-reset relay for bridge balance and relay for excitation on/off). Each requires 10 mA at 5 Vdc, except excitation on/off 25 mA.

 

Filter:
  • Characteristic: Low-pass active 2-pole Butterworth standard.
  • Frequencies (-3 + 1 dB): 10, 100, 1000 and 10 000 Hz and wide-band.
  • Outputs Filtered: Any 1 or 2 or all (switch-selected on PC board).

 

Amplifier Outputs:
  • Standard Output: + 10V at 5 mA, min.
  • Tape Output: + 1.414V (1 Vrms) at 5 mA, min.
  • Galvanometer Output: + 10V at 75 mA, min, current-limited at 100 mA, max (minimum load resistance for 0.05% linearity: 50Ω).
  • Galvanometer attenuator (0-100%) and zero adjust (+ 1V) on front panel.
  • Linearity at dc: 0.02%.
  • Any output can be short-circuited with no effect on others.

 

Playback:
  • Input: + 1.414V full scale; input impedance 20 kil ohms.
  • Gain: X1 to tape output; X7.07 to standard output.
  • Filter Selection: As specified above.
  • Outputs: All three as specified above.

 

Power:
  • 105 to 125V or 210 to 250V (switch-selected), 50/60 Hz, 10 watts, max.

Signal Conditioning Amplifier System:

While the Model 2311 provides wider frequency response and more versatile excitation, both 2310 and 2311 Signal Conditioner/Amplifier Modules accept inputs from strain gages, load/pressure/dc displacement transducers, potentiometers, RTD's and nickel temperature sensors, without any internal modification.


Specifications:

All specifications nominal or typical at +23 deg C unless noted.

Input:
  • Strain Gages: Quarter(120Ω or 1000Ω and 350Ω), half or full bridge (50 to 1000Ω). Dummy resistors installed.
  • Transducers: Foil or piezoresistive strain gage types.
  • Potentiometer: DCDT displacement transducer.

 

Excitation:
  • 12 settings: 0.5, 0.7, 1, 1.4, 2, 2.7, 3.5, 5, 7, 10, 12 and 15 Vdc + 1%, max.
  • Current: 0-100 mA, limited at 175 mA, max.
  • Regulation (0-100 mA, + 10% line change): + 0.5 mV + 0.04%, max measured at remote sense point. (Local sense: -5 mV, typical, at 100 mA, measured at plug.)
  • Remote Sense Error: 0.0005%/Ω of lead resistance (350Ω load).
  • Noise and Ripple: 0.05% p-p, max (dc to 10 kHz).
  • Stability: 0.02%/deg C.
  • Level: Normally symmetrical about ground; either side can be grounded with no effect on performance.

 

Bridge Balance:
  • Method: Counter-emf injection at preamp; automatic electronic; dual range; can be disabled on front panel.
  • Ranges (auto ranging):
    + 5000 microstrain (1% bridge unbalance or 2.5 mV/V), resolution 2.5 microstrain (0.0012 mV/V).
    + 25 000 microstrain (5% bridge unbalance or 12.5 mV/V), resolution 12.5 microstrain (0.006 mV/V).
  • Balance Time: 2 seconds, typical.
  • Manual Vernier Balance: + 50 microstrain ( + 0.025 mV/V).
  • Interaction: Essentially independent of excitation and amplifier gain.
  • Storage: Digital; up to 2 years without line power.

 

Amplifier:
  • Gain: 1 to 11 000 continuously variable. Direct-reading.
          + 1% max of reading; + 0.5% max of full scale vernier setting.
  • Ten-turn counting knob (X1 to X11) plus decade multiplier (X1 to X1000).
  • Frequency Response (all gains >5, full output):
          dc coupled: dc to 25 kHz, -0.5 dB max.
                            dc to 65 kHz, -3 dB (typical at 40% output).
          ac coupled: 5 Hz to 25 kHz, -0.5 dB.
  • Input Impedance: 100 MΩ, min, differential or common-mode, including bridge balance circuit.
  • Bias Current: + 50 nA, typical each input.
  • Source Impedance: 0 to 1000Ω each input.
  • Common-Mode Voltage: + 10V.
  • Common-Mode Rejection (gain over X100):
    Shorted input: 100 dB, min, at dc; 90 dB, min, dc to 1 kHz.
    350Ω balanced input: 90 dB, typical, at 1 kHz.
  • Stability (gain over X100): + 2 microvolts/deg C, max, referred to input (RTI).
  • Noise (gain over X100, all outputs):
    0.01 to 10 Hz: 1 microvolt p-p RTI.
    0.5 Hz to 125 kHz: 5 microvolts rms, max, RTI


* Can be used with a Vishay 2360 4 Bay Portable Adapter