Channel number of HV |
4 |
Impulse voltage |
100V~5000V, 10V step,5%±10V |
Inductance test range |
≥20uH |
Impulse energy |
Max.: 0.25 Joule |
Test speed |
Single-winding: Max. 12 meas/sec, 4-winding: Max. 2 meas/sec |
Applied pulses |
Max. 32 of test pulse; max.8 of demagnetizing pulse |
Input impedance |
10M |
Display |
800x480 dot 65k color TFT, waveform display range: 650x256, supports 1.5 times magnifying display |
Waveform acquisition |
Sampling rate: max. 100Msps, adjustable 10 levels; resolution: 8 Bits ; memory depth: 6500 bytes; Sample average: 1-32 |
Comparison methods |
Area, differential area, corona and differential phase |
Test repeated accuracy |
±1% |
Comparison test |
yes |
Waveform measurement |
Voltage/frequency/time of the waveform |
Trigger Mode |
Manual (including foot switch)/External/Internal/Bus |
Detection Output |
OK/NG display, LED / alarm |
Test statistics |
Statistics function of measurement time and results, up to 20000 pieces |
Internal Memory |
120 groups |
U-disk Memory |
600 groups (standard waveform data and test setup parameters) |
HANDLER signal |
START, STOP, PASS, FAIL, BUSY, EOC |
Interface |
USB Slave, RS232C, IEEE488(option); USB Host(support storage of BMP, GIF, PNG image files and firmware upgrade) |
Power supply |
220V±10% 50Hz±5% |
Power consumption |
≤50VA |
Weight |
8.0Kg |
Dimensions (WxHxD) |
400mm x 132mm x 350mm |
Temperature |
0℃ - 40℃ |
Humidity |
≤75% R.H. |