IEC Standard Test Mode |
Contact discharge mode : 2.0kV, 4.0kV, 6.0kV and 8.0kV step
Air discharge mode : 2.0kV, 4.0kV, 8.0kV and 15.0kV step |
Manual Test Mode |
Contact / Air discharge mode, Arbitrary setting during 0.2kV - 30.0kV
Sweeping function built-in, recordable up to 99 units |
Sequence Test Mode |
Enables to operate units set in MANUAL mode continuously
Max. 22 steps / 1 program, recordable up to 20 |
Discharge Repetition Cycle |
0.05s - 99.9s±10% / Manual
Set step : 0.05s (0.05 - 10.0s), 0.10s (10.0 - 99.9s) |
Number of Time of Discharge |
1 - 60,000 times, Preset 1 time step or continuous preset |
Default R/C Network |
150 pF / 330 Ω |
Voltage Resolution |
10 V steps |
Discharge Polarity |
Positive & Negative |
Operating Modes |
Continuous, Single & Random |
Weight |
8.5 kg |