Chroma 63630-80-60 DC Electronic Load 80 V, 60 A, 300 W

Rent Chroma 63630-80-60 DC Electronic Load 80 Volts, 60 Amps, 300 Watts
Enlarge Image
Chroma 63630-80-60
Request Quote Print Page
Datasheet
The Chroma 63630-80-60 DC Electronic Load 80 Volts, 60 Amps, 300 Watts is designed for testing multi-output AC/DC power supplies, DC/DC converters, chargers, batteries, adapters, and power electronic components. They are excellent for research, development, production, and incoming inspection applications. The 63630-80-60's state of the art design uses DSP technology to simulate non-linear loads using a unique CZ operation mode allowing realistic loading behavior. The 63630-80-60 can draw its rated current under very low voltage (0.4V typical). This unique feature guarantees the best loading performance for modern Point-of-Load conditions and fuel cells. The 63630-80-60 can simulate a wide range of dynamic loading applications, with programmable load levels, slew rates, duration, and conducting voltage.
 
Chroma 63630-80-60 Features
  • Up to 10 channels in one mainframe,fit for testing multiple output SMPS
  • 0.4V @ 80A (Typical) low voltage operating characteristics
  • Flexible CC, CR, CV and CP operation modes
  • CZ mode for turn on capacitive load simulation
  • Parallel mode for high current and power application up to 2kW
  • Multi channel synchronous control
  • Auto frequency sweep up to 50kHz
  • Real time power supply load transient response simulation and Vpk+/- measurement
  • User programmable 100 sequential front panel input status for user-friendly operation
  • Precision voltage and current measurement
  • Precision high speed digitizing measurement/ data capture
  • Voltage, current and Pmax measurement for OCP/OLP testing
  • Timing measurement for batteries
  • Short circuit simulation
  • Self-test at power-on
  • Full protection : OC, OP, OT protection and OV alarm
  • Ethernet, USB and GPIB interfaces
Chroma 63630-80-60 Specs
Constant Current Mode
Range: 0~0.6A 0~6A 0~60A
Resolution: 0.01mA 0.1mA 1mA
Accuracy: 0.1%+0.1%F.S.
Constant Resistance Mode
Range: CRL : 0.015~30Ω (300W/6V)
CRM: 0.3~600Ω (300W/16V)
CRH : 1.5~3kΩ (300W80V)
Resolution: 0.9864mS
Accuracy: 0.1%+0.2S (6V)
0.1%+0.03S (16V)
0.1%+0.01S (80V)
Constant Voltage Mode
Range: 0~6V 0~16V 0~80V
Resolution: 0.1mV 1mV 1mV
Accuracy: 0.05%+0.1%F.S.
Constant Power Mode
Range: 0~6W 0~30W 0~300W
Resolution: 3.2mW 32mW 320mW
Accuracy: 0.3%+0.3%F.S.