Test Level |
Test voltage |
200V - 30,000V ± 5% for contact discharge
200V - 30,000V ± 5% for air discharge |
Resolution |
100V steps |
Holding time (AD) |
More than 5s |
Mainframe |
HV power supply |
Up to 30kV DC |
Power supply |
88V - 250V AC
12V - 18V DC |
Battery |
Approx. 1 hour operation; Buffer battery is automatically recharged when connected to AC mains supply
|
Interface |
USB to control the ESD generator and to download pre-programmed test routines from esd.control
|
Output |
Contact Discharge Pulse |
As per EN/IEC 61000-4-2; 150pF/330ohm |
Rise time |
As per EN/IEC 61000-4-2 Ed.2 0.8ns +/- 25% |
First peak current value |
As per EN/IEC 61000-4-2 Ed.2
7.5A ± 15% @2kV
15A ± 15% @4kV
15A ± 15% @4kV
15A ± 15% @4kV |
Current value at 30ns |
As per EN/IEC 61000-4-2 Ed.2
4.0A ± 30% @2kV
8.0A ± 30% @4kV
12.0A ± 30% @6kV
16.0A ± 30% @8kV |
Current value at 60ns |
As per EN/IEC 61000-4-2 Ed.2
2.0A ± 30% @2kV
4.0A ± 30% @4kV
6.0A ± 30% @6kV
8.0A ± 30% @8kV |
Weight |
Approx. 5.1kg (ESD 30N control unit) |
Supply voltage |
AC: 88V - 264V, 50/60Hz
DC: 11V - 16VDC |