Particle Counting |
Size Range |
D.100 to 10.0 µm |
Partide Channel Sizes |
9110·01: 0.10,D.lS,0.2,0.2S,0.3, 05,l.O,s.oµm |
Size Resolution |
<1S% @ 0.2 µm (per ISO 21S01 4) |
Counting Efficiency |
SO% at D.100 µm;100% for partides > 0.15µm
(per ISO 21S01 4 and]IS) |
Concentration Limit |
100,000 particles/ft3(3,SOO,OOO/m3)
@ 10%coincidence loss |
Light Source |
Enhancedactive cavity HeNelaser (4 year warranty) |
Zero Count |
< 1count per S minutes (per ISO 21S01·4 and ]IS) |
Flow Rate |
1CFM (28.3Umin) ±S% accuracy (meets ISO 21S01·4 and ]IS B9921) |
Flow Rate Control |
Electronic. automatic closed loop (patented"flow control technology) |
Calibration |
NIST traceableusing TSI calibration system |
Calibration Frequency |
Recommended minimum of once per year |
Sampling Modes |
Manual, automatic. beep;cumulative/ differential; count or concentration |
Sampling Time |
1second to 99 hours |
Sampling Frequency |
1to 9999 cyclesor continuous |
Exhaust |
Internally HEPA filtered |
Communication Mode |
Modbus<& TCP over Ethernet or USB |
Audible Alarm |
Built-in; >BS dB at 1meter (adjustable) |
Status Indicator |
Low battery,flow,laser |
Alarm Output |
Dry contacts, normally closed when alarm is engaged |
Alarm Limits |
Programmablefor all particlechannels |
Display |
VGA S.7 inch (14.S·cm) touch screen display |
Languages |
English, German, French, Spanish,Japanese, Chinese (simplifiedi Italian |
Software |
Compatible with TrakPro Lite Secure and FMSs
software |
Printer |
Built-in thermal printer (also availablewithout printer) |
Printer Output |
Prints in allavailable languages |