The Lighthouse SOLAIR 1100LD Airborne Particle Counter is a cutting-edge device with a size range of 0.10 to 1.0 µm and flow rate of 1.0 CFM (2.83 LPM). With data storage of up to 3,000 sample records and 5.7 inch color display the SOLAIR 1100LD is the ideal model for easy operation and user-friendly applications. The Lighthouse SOLAIR 1100LD Airborne Particle Counter has a variety of language options, including English, French, German, Spanish, and more.
Features
- 0.1- 1.0 µm sensitivity
- 1.0 CFM (28.3 LPM)
- Up to 8 channels of simultaneous count data
- 21 CFR Part 11 compliant secure USB flash transfer
- Laser Diode Sensor Technology
- 5.7 inch color touch screen display
- Thermal printer
- Stainless steel construction
- Ethernet/RS485/Modbus, 4x 4-20mA analog outputs
- 3,000 record data storage
- Report wizard: ISO-14644-1, EU GMP Annex 1, FS-209E
Specifications
Model |
1100LD |
Size range |
0.10 - 1.0 µm |
Counting efficiency |
50% @ 0.1 µm
100% for particles >0.15 µm per ISO 21501-4 |
Flow rate |
1.0 CFM (2.83 LPM) |
Laser source |
Laser Diode |
Zero count level |
<1 count / 5 minutes per ISO 21501-4 |
Data storage |
3,000 sample records
Rotating buffer
Particle and environmental data
Location and time |
Power |
100 - 240 VAC
50 - 60 Hz
External AC/DC adapter |
Operating temperature |
50°F to 104°F (10°C to 40°C) |
Storage temperature |
14°F to 122°F (-10°C to 50°C) |
Dimensions
(L x W x H) |
14.65 x 8.0 x 10.4 in (37 x 20 x 26 cm) |
Weight |
With battery: 23.35 lb. (10.6 kg)
Without battery: 20.25 lb. (9.19 kg) |