Amplifier Research 1000L RF Amplifier 10 kHz - 220 MHz

Amplifier Research 1000L RF Amplifier, 10 kHz - 220 MHz
Enlarge Image
Amplifier Research 1000L
Request Quote Print Page
Datasheet

The Amplifier Research 1000L is an economical, self-contained, air-cooled broadband amplifier designed for laboratory applications that require instantaneous bandwidth, high gain and high power output. Housed in a stylish contemporary enclosure, the 1000L is smaller than competitive units with similar power levels. All operating controls are functionally grouped on the front panel for simplicity of operation. These include modern, lighted push-button switches for the command functions, POWER, STANDBY, OPERATE and PULSE, a control for setting the output level of the amplifier, and a meter for monitoring critical operating voltages and currents. Remote control is provided through a rear panel mounted connector. A highly versatile unit, the 1000L features rugged circuitry and a quick-acting, solid state crowbar circuit to protect the final amplifier tubes from damage due to internal arcing. An electronic circuit is provided to enable rapid gating or blanking of the amplifier.

 

Specifications:

Power Output
High Range Pulse: 2500 Watts to 150 MHz (Minimum) / 1750 watts to 220MHz
Duty Cycle: 15%
Pulse Width: 8 milliseconds
CW: 1200 Watts (minimum)
Low Range: 100 Watts nominal
Flatness (high range) ±1.5 dB
Frequency Response 10 kHz - 220 MHz instantaneously
Input For Rated Output 1.0 milliwatt maximum
Gain (at maximum setting) High Range 61 dB minimum
Low Range 47 dB minimum
Gain Adjustment (continuous range) 18 dB minimum
Input Impedance 50 ohms, VSWR 1.5:1 maximum
Output Impedance 50 ohms, nominal
Mismatch Tolerance 100% of rated power without foldback. Will operate without damage, or oscillation with any magnitude and phase of source and load impedance.
Modulation Capability Linear amplitude and phase response to over 80 MHz allows faithful reproduction of AM, FM, Pulse, or phase modulation appearing on the input signal
Harmonic Distortion at 750 Watts Above 120 MHz: Minus 30 dBc maximum
Below 120 MHz: Minus 15 dBc maximum / Minus 18 dBc nominal
Third Order Intercept Point 66 dBm Typical
Gating Characteristics Pulse Mode Pedestal/CW Mode Blanking Signal (into 180 ohms): Plus or minus 2.5 VDC to 6.0 VDC
Rise time: 20 microseconds maximum
Fall time: 4 microseconds maximum
RF Rise/Fall Time: 10 nanoseconds maximum
RF Pulse Droop: 1.0% maximum at 8 milliseconds
Primary Power (specify one) 200/208 ±5% VAC, 3 phase, 50/60 Hz
380/415 ±5% VAC, 3 phase, 50/60 Hz
400/415 ±5% VAC, 3 phase, 50/60Hz
15.2 kVA nominal
Cooling Forced air (self contained fans)
Weight 239 kg (525 lb)
Size (W x H x D) 22.1 in x 59.0 in x 23.0 in (56.1cm x 149.9cm x 58.4 cm)