16A Programmable Step Transformer for NSG3040/NSG3060 units for 61000-4-11 requirements
The automatic step transformer type INA 6502 is a standard accessory for the Teseq Modula 6000 instrumentation series. It enables the ac power supply voltage, which is fed to the EUT, to be set to 0%, 40%, 70% or 80% of UIN under program control.
The INA 6502 complies with the requirements of IEC/EN 61000-4-11(2004) for voltage dips and under-voltages. A large number of product standard specifications can be fully met as they do not ask for continuous and arbitrary voltage variations.
Features
- Integrated function unit to Modular system
- Power line dips to 0%, 40%, 70% and 80%
- Complies with EN/IEC 61000-4-11
Specifications
Input voltage UIN: | 0 to 250 VAC,50/60 Hz |
Output voltage UVAR: | 4 steps:0% - 40% - 70% - 80% of UIN |
Accuracy: | ±5% |
Connectors: | Harting type HAN3A, compatible with Modula safe power linking concept |
Instrument supply: | Selectable 100–110 VAC, 220–250 VAC, 15 VA, 47 to 63 Hz |
Weight: | 15 kg (33 lbs) approx. |