Haefely PLINE 1610 Voltage / Interruption Tester for IEC 61000-4-11

Rent, lease, rent to own Tester For Voltage Dips, Interruptions and Variations for IEC 6100-4-11
Enlarge Image
Haefely PLINE 1610
Request Quote Print Page
Datasheet Brochure

Voltage dips and short interruptions are caused by faults in the network or by sudden large load changes. Two or more consecutive dips or interruptions may occur due to changes in the load with time. Voltage variations are caused by continuously varying loads connected to the network.
 

The Haefely PLINE-1610 Voltage, Interruption and Variation Tester contains all the features expected from a top quality generator. A reliable semiconductor switch is built-in the unit. The semiconductor switch operates to generate programmed Dips and Interrupts with the full 500 A in-rush current capability and provide increased current (in the case of a constant power load) up to 40 A. This totally integrated solution guarantees compliance with standard requirements in particular the interrupt rise and fall time and inrush current.
 

The Haefely PLINE1610 has two integrated variable transformers (variacs). One transformer is used to adjust the nominal voltage in the range 0 to (Vin+10%) up to maximum 250 V. A second transformer responds to programmed inputs and adjusts the Dip voltage or slow variation voltage. The Haefely MAG 100 for EN/IEC 61000-4-8 generates power frequency magnetic fields up to 100 A/m at 50/60 Hz.

 
Haefely PLINE 1610 Features
  • Totally reproducible test results
  • Fully automated, time saving design
  • Integration in a full EMC test system
  • a compact test system for IEC 61000-4-11, EN 61000-4-11
  • user friendly test programming
  • no additional voltage sources wiring or variacs are required
Haefely PLINE 1610 Specs
Power unit
Output voltage range 0 to 264 VAC
Frequency range 48 Hz - 62 Hz
Output current up to 16 A rms (40 A for 5 s)
Inrush current >500 A at 230 V
In Acc. EN/IEC61000-4-11 >250 A at 115 V
Irms and Urms measurement 3%
Dips/Interruption amplitude 0 to 80%
Synchronization asynchronous
synchronous 0 to 360°
resolution 1°
Dips/Interruption duration 30 μs to 70 min.
Dips/Interruption repetition rate 3 ms to 11 hours
Test time 1 s to 100 hours
Voltage variation amplitude 0 to 115%
Voltage variation time 2 s to 10 hours
Control unit
Memory for test sequences 18 program and 18 test memory locations
EUT current limit parameter software controlled 0 to 16 A
Linear transitions/ramps voltage, phase, interruption and repetition time
BNC-monitor output voltage 40:1 and current 4 A/V
Trigger manual or automatic
Remote control interface serial (RS232) as standard
Optional interfaces IEEE 488 (GPIB)
Printer interfaces Centronics
Haefely PLINE 1610 Standards