Capacitive Coupling Clamp |
Oscillation Frequency |
1, 10, 30 MHz |
Usable Cable Diameter |
4 – 70 mm |
Maximum Insulation Voltage |
5 kV (1.2 / 50 us) |
Data Line Common Specifications |
Maximum EUT Voltage AC |
300 V |
Maximum EUT Voltage DC |
200 V |
Maximum EUT Current |
3 A per line |
Coupling Capacitance Slow DOW |
0.5 uF |
Coupling Modes |
Common mode
Differential mode |
Damped Oscillatory Waves (SLOW) |
Oscillation Frequency |
100 kHz & 1 MHz |
Voltage Range |
Up to 4.4 kV |
Source Impedance |
200 Ω |
Burst Repetition at 100 kHz |
Up to 50 Hz |
Burst Repetition at 1 MHz |
Up to 500 Hz |
Damped Oscillatory Waves (FAST) |
Oscillation Frequency (MHz) |
3, 10, 30 |
Voltage Range |
Up to 4.4 kV |
Source Impedance |
50 ohms |
Burst Repetition All Frequencies |
Up to 5 kHz |
Surge IEC 60255-27 0.5J 500 Ohm |
Waveform at No Load |
1.2 / 50 us |
Rise Time |
1.2 μs |
Adjustable Voltage Range |
500V – 8000 V |
Damped Oscillatory Waves Magnetic Field with MF1000-1 |
Oscillation Frequency |
100 kHz & 1 MHz |
Antenna Dimension |
1 m x 1 m |
Current Range |
1 – 150 A (100 A/m) |