Olympus 45MG Ultrasonic Thickness Gage

Rent Olympus 45MG Ultrasonic Thickness Gage
Enlarge Image
Olympus 45MG
Request Quote Print Page
Manual Datasheet
The Olympus 45MG Ultrasonic Thickness Gage is an advanced ultrasonic thickness gauge packed with standard measurement features and software options. This unique instrument is compatible with the complete range of Olympus dual element and single element thickness gauge transducers, making this innovative instrument an all-in-one solution for virtually every thickness gauge application.

In its basic configuration the 45MG is a simple and straightforward gauge that requires minimal operator training to tackle most common thickness gauging applications. With additional optional software options and transducers however, the 45MG can become significantly more advanced and take on applications well beyond that of a typical entry-level gauge. Furthermore, most options are available individually at the time of purchase or can be added in the future as your needs change.
 
Olympus 45MG Features
  • Compatible with full line of Olympus dual element transducers for thickness measurements on internally corroded metals
  • Min./Max. mode
  • Two Alarm modes
  • Differential mode
  • Time-based B-scan
  • Reduction Rate
  • Gain Adjust (standard, high, and low)
  • Password instrument lock
Olympus 45MG Specs
Measurements
Dual element transducer measurement mode Time interval from a precision delay after the excitation pulse to the first echo
Echo-to-Echo (optional) Time interval between two successive back-wall echoes to eliminate paint or coating thickness
THRU-COAT® measurement (optional) Measurement of true metal and coating thicknesses with a single back-wall echo (with D7906-SM, D7906-RM, and D7908 transducers)
Single element transducer measurement modes (optional) Mode 1: Time interval between the excitation pulse and the first back-wall echo
Mode 2: Time interval between the delay line echo and the first back-wall echo (with delay or immersion transducers)
Mode 3: Time interval between successive back-wall echoes following the first interface echo after the excitation pulse (with delay line or immersion transducers)
Thickness range 0.080 mm to 635 mm (0.003 in. to 25.0 in.) depending on material, transducer, surface conditions temperature, and selected configuration (Full range requires single element option)
Material velocity range 0.508 mm/μs to18.699 mm/μs (0.020 in./μs to 0.7362 in./μs)
Resolution (selectable) Low: 0.1 mm (0.01 in.)
Standard: 0.01 mm (0.001 in.)
Single Element option: 0.001 mm (0.0001 in.)
Transducer frequency range Standard: 2.25 MHz to 30 MHz (-3 dB) High Penetration (Single Element option): 0.50 MHz to 30 MHz (-3 dB)