Pressure Ranges |
P3830 |
500 psi to 30 000 psi, or 40 bar to 2 000 bar, or 4 MPa to 200 MPa |
Accuracy |
Standard accuracy is 0.02 % of Reading*. Optional accuracy of 0.015% of Reading is available. ISO/IEC
17025 accredited calibration certificate is standard
*Accuracy is based on % of Reading from 10% to 100% of the piston range when used in accordance with
the corrections found on the calibration certificate. Below 10%, ± (accuracy class) x 10% of the piston range. |
Materials of Construction |
Standard weight material |
Series 3 non-magnetic, austenitic, stainless steel
Density: 7.8 g/cP3 |
Piston material |
Tungsten carbide with nickel binder |
Cylinder material |
Tungsten carbide with cobalt binder |
Thermal coefficient of
expansion |
11 ppm/°C |
General |
Test port adapters |
3/4-16 UNF; 9/16 in – 18 UNF (Autoclave); 3/8 in BSP; 1/2 in BSP and 3/4 in BSP |
Weight |
30 Kg (66 lb) Instrument base only |
Dimensions (W x D x H) |
455 mm X 340 mm X 478 mm ( 18 in X 13.5 in X 19 in) |
Reservoir volume |
235 cm3 (14.3 in3) |
Intensifier ratio |
7 to 1 |
Seal materials |
Nitrile |
Operating fluid |
Sebacate, our reference 3069551 |