Principle of Analysis |
Water-based condensation particle counter |
Range |
0.001 to 106 particles/cm3 |
Particle Size Range (D50) Range |
0.007µm (7nm) to ~3 µm |
Particle Concentration Accuracy |
+/-10% at 106 particles/cm3 |
Measurement Averaging Time |
1 to 3600 seconds (1 hour) |
Response Time |
<3 seconds to 95% in response to concentration step change (high flow mode: 3 lpm);
<5 seconds to 95% in response to concentration step change (low flow mode: 0.6 lpm) |
Flow Rate |
0.6 +/- 0.06 lpm (low flow mode);
3.0 +/- 0.3 lpm (high flow mode) |
Environmental Conditions |
10 – 38ºC (50 – 100ºF) non-condensing; (Recommended: 20 – 25ºC) |
Power Requirements |
100-240 VAC, 50/60HZ, 145W max |
Analog Output Ranges |
0-4V proportional to log concentration |
User Interface |
Color touchscreen display |
Communications |
Ethernet, RS232 and USB |
Dimensions (HxWxD) |
8” x 19” x 12” (203 x 483 x 305 mm) |
Weight |
22 lbs (9.9kg) |