Thermocouple Measurement and Source Accuracy |
Measure and Simulate |
Standard |
Temperature Range (oC) |
Accuracy (oC) |
S |
IEC 584 |
-50 to 1768 |
Measure: 1, 0.6, 0.7
Source: 1.1, 0.6, 0.8 |
R |
IEC 584 |
-50 to 1768 |
Measure: 1.4, 0.6, 0.6
Source: 1.4, 0.6, 0.7 |
B |
IEC 584 |
0 to 1820 |
Measure: 3.8, 0.9, 0.6
Source: 3.8, 0.9, 0.7 |
K |
IEC 584 |
-270 to 1372 |
Measure: 1, 0.4, 0.3, 0.4
Source: 1.1, 0.5, 0.3, 0.5 |
N |
IEC 584 |
-270 to 1300 |
Measure: 1.5, 0.5, 0.4
Source: 1.6, 0.6, 0.5 |
E |
IEC 584 |
-270 to 1000 |
Measure: 0.6, 0.3, 0.2, 0.2, 0.,2
Source: 0.7, 0.3, 0.2, 0.3, 0.4 |
J |
IEC 584 |
-270 to 1200 |
Measure: 0.3, 0.3
Source: 0.3, 0.4 |
T |
IEC 584 |
-270 to 400 |
Measure: 0.8, 0.4, 0.2
Source:0.9, 0.4, 0.2 |
C |
ASTM E988 |
0 to 2315 |
Measure: 0.5, 0.7, 1
Source: 0.5, 0.9, 1.4 |
D |
ASTM E988 |
0 to 2320 |
Measure: 0.5, 0.4, 0.6, 0.9
Source: 0.5, 0.5, 0.9, 1.3 |
G |
ASTM E1751 |
0 to 2315 |
Measure: 2.4, 0.5, 0.4, 0.7
Source: 2.4, 0.5, 0.5, 1 |
L |
DIN43710 |
-200 to 900 |
Measure: 0.2, 0.2, 0.2
Source: 0.3, 0.2, 0.3 |
U |
DIN43710 |
-200 to 600 |
Measure: 0.4, 0.2
Source: 0.4, 0.3 |