Milliamp and voltage |
Source 0.00 - 24.00 mA, 0.00 - 10.25 VDC and 0.00 - 80.00 mV
Read to 24.00 mA, 00.00 - 10.,25 VDC, 0.0 - 60.0 VDC, 00.00 - 80.00 mV
Simulate 2 Wire transmitters
Power up transmitters and loops with built-in 24 V power supply
Simplify HART hook ups with built in 250 ohm resistor |
Operating temperature |
-20 to 60 °C (-5 to 140 °F) |
Source V DC range and accuracy |
-10.00 to 80.00 mV, ≤ ±(0.03% of full scale)
0.00 to 10.25 V, ≤ ±(0.03% of full scale +0.005 V) |
pH source |
Range and resolution: 0.000 to 14.000 p
Accuracy in mV: ≤ ±(0.02% of Reading in mV + 0.1 mV)
Accuracy in pH: ≤ ±0.003 pH @ 25°C |
Thermocouple source |
Accuracy: ≤ ±(0.02% of Reading + 0.01 mV)
Output impedance: <1 ohm |
Frequency range |
1 - 2000 CPM, 0.01 - 999.99 Hz, 0.1 - 9999.9 Hz,
0.001 - 20.000 kHz |
Dimensions
(L x W x H) |
5.63 x 3.00 x 1.60 in
143 x 76 x 41 mm |
Weight |
12.1 oz (0.34 kg) |