Fluke 725 Multifunction Process Calibrator

Fluke 725 Multifunction Process Calibrator
Enlarge Image
Fluke Calibration 725
Request Quote Print Page
Manual Other1

Take along the new Fluke 725 Multifunction Process Calibrator, and you're equipped to test and calibrate almost any process parameter. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.
 

Do you need to calibrate transmitters? The 725's split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725 has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.
 

From the moment you pick it up, the Fluke 725 is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.
 


DC Voltage Measurement
 

  • Range: 30V, 20 V, 90 mV
  • Resolution: 0.001V, 0.001 V, 0.01 mV
  • Accuracy, (% of Reading + Counts): 0.02 % + 2


Voltage Source
 

  • Range: 100 mV, 10 V
  • Resolution: 0.01 mV, 0.001 V
  • Accuracy, (% of Reading + Counts): 0.02 % + 2
  • Maximum load: 1 mA


Millivolt Measurement and Source
 

  • Range: -10 mV to 75 mV
  • Resolution: 0.01 mV
  • Accuracy: ±(0.025 % + 1 count)
  • Maximum input voltage: 30 V


DC mA Measurement and Source
 

  • Range: 24 mA
  • Resolution: 0.001 mA
  • Accuracy, (% of Reading + Counts):  0.02 % + 2
  • Drive capability: 1000W at 20 mA

Ohms Measurement
  • ohms Range: 0 to 400 W, 400 to 1.5 kW, 1.5 to 3.2 kW
  • 4-Wire:  0.1, 0.5, 1±W
  • 2- and 3-Wire: 0.15, 1.0, 1.5 ±W
  • Excitation Current: 0.2 mA
  • Maximum input voltage: 30 V


Ohms Source
 

  • ohms Range: 15 to 400 W
    • Excitation Current from Measurement Device: 0.15 to 0.5 mA
    • Accuracy: ±0.15W
  • ohms Range: 15 to 400 W
    • Excitation Current from Measurement Device: 0.5 to 2 mA
    • Accuracy: ±0.1W
  • ohms Range: 400 to 1.5 kW
    • Excitation Current from Measurement Device: 0.05 to 0.8 mA
    • Accuracy: ±0.5W
  • ohms Range: 1.5 to 3.2 kW
    • Excitation Current from Measurement Device: 0.05 to 0.4 mA
    • Accuracy: ±1W
  • Resolution:
    • 15 to 400 W 
    • 400 to 3.2 kW


Frequency Measurement
 

  • Range: 2.0 to 1000.0 CPM
    • Resolution: 0.1 CPM
    • Accuracy: ± (0.05 % + 1 count)
  • Range: 1 to 1100
    • Resolution: 0.1 Hz 
    • Accuracy: ± (0.05 % + 1 count)
  • Range: 1.0 to 10.0 kHz
    • Resolution: 0.01 kHz
    • Accuracy: ± (0.05 % + 1 count)
  • Sensitivity: 1V peak-to-peak minimum
  • Waveform: squarewave