Amplitude Accuracy (field aligned with sensor axes): |
Without correction factors applied: ±1.0 dB @ 10 MHz; with correction factors applied: Typical expanded measurement uncertainty (95% confidence interval), 0.8 dB, 100 kHz – 1 GHz; 1.4 dB, 1GHz – 6 GHz |
Response Time/Sampling Rate (through FL7000): |
20 msec/up to 50 samples per second, USB and GPIB only |
Isotropic Deviation (measured at the ortho angle): |
±0.5 dB @ 10 MHz
±0.5 dB, 0.5 MHz – 2 GHz (typical) |
Operating Range: |
0.5 – 800 V/m, 100 kHz – 1 GHz
0.5 – 600 V/m, 1 GHz – 4 GHz
0.7 – 800 V/m, 4 GHz – 6 GHz |
Linearity, 0.5 to 800 V/m: |
±0.5 dB and ±0.3 V/m |
Power Requiremens |
Laser powered from FL7000 interface |