Keithley 2290-5 HV DC Power Supply | 5 kV, 5 mA
The Keithley 2290-5 High Voltage Power Supply supports testing in high voltage applications and physics research with up to 5 kV output and 5mA current capacity. The model accurately measures voltage and current, displaying detailed settings for control. Users can adjust settings manually or program them via the IEEE-4888 interface or analog voltage. Reach out today to learn more about the 2290-5.
Voltage Range | ||
Output Voltage | Maximum Output Current | Conditions |
50 – 5000 VDC | 5.000 mA DC | No filter |
50 – 3000 VDC | 5.000 mA DC | Filter 1 |
50 – 5000 VDC | 3.000 mA DC | Filter 2 |
Voltage Set Accuracy | ±(0.01% of setting + 2.5 V) | |
Voltage Display Accuracy | Voltage set accuracy ±1 V, typical (±2 V, maximum) | |
Voltage Resolution | 1 V (set and display) | |
Voltage Limit Range | 0 – 100% full scale | |
Voltage Regulation | Line: 0.001% for ±10% line voltage change Load: 0.005% for 100% load change, typical |
|
Output Ripple | 0.002% of full scale, V RMS, maximum - No filter 1.0 mV RMS @ 1 kV – Filter 1 or Filter 2 2.0 mV RMS @ 3 kV – Filter 1 or Filter 2 3.0 mV RMS @ 5 kV – Filter 2 |
|
Rise Time | 1.5 seconds for 0 V setting to within 1 V of 5000 VDC – No filter 3 seconds for 0 V setting to within 1 V of 3000 VDC – Filter 1 4 seconds for 0 V setting to within 1 V of 5000 VDC – Filter 2 |
- 5 kV Source voltage
- 1 μA current measurement resolution
- Low noise for precision sourcing and sensitive measurements
- Safety interlock controls high voltage output
- IEEE-488 programmable
- Protection module prevents damage to low voltage instrumentation
- High voltage device breakdown testing
- High voltage component testing
- Insulation testing
- High voltage resistivity measurements
- High energy physics research