Dimensions (L x W x H) |
17.6 in x 13 in x 5.9 in (446 mm x 330 mm x 149 mm) |
Weight |
30 lb. (14 kg) |
AC Hipot |
Voltage Range: 100 to 5000 VAC RMS
Voltage Accuracy: 1% of reading + 5 V
Max Output Current: 20 mA for 60 s, 10 mA continuous
Min/Max Current Limits: 0.01 to 20 mA
Current Accuracy: 1% of reading + 0.05 mA |
DC Hipot |
Voltage Range: 100 to 6000 VDC
Voltage Accuracy: 1% of reading + 5 V
Max Output Current: 7.5 mA
Min/Max Current Limits: 0.01 to 7.5 mA
Current Accuracy: 1% of reading + 0.05 mA |
Insulation Resistance |
DC Output Voltage: 500 V or 1000 V
Resistance Range: 1 to 9999 MΩ |
Continuity Measurement |
Test Current: 100 mA ± 10 mA
Resistance Range: 0.001 to 1.500 Ω |
Operating Temperature |
0°C to 40°C, 80% RH max |
Storage Temperature |
-10°C to 70°C, 70% RH max |
Accuracy |
Stated specifications apply for 1 year at 23°C ± 5°C |
Model |
V53 |