The Vitrek V63 AC/DC/IR Electrical Safety Analyzer is part of the V60 Series of electrical safety testers. And when it comes to getting the job done, the Vitrek V63 comes loaded with standard features that help you automate the task at hand. With the built-in RS232 interface, you can download and run test sequences, as well as, retrieve test results and data.

The Vitrek V63 has been engineered and manufactured to meet the most demanding electrical product safety test requirements. So, whether your job is on an automated production line or in a product safety engineering lab, the V63 packs the performance you need to get the job done right.
 
Vitrek V63 Specs
AC Hipot
Voltage Range 0.100kV to 5.000kV
Voltage Setting 5 Volt Increments
Voltage Regulation 1% + 5V
(Line & Load)
Voltage Accuracy 1.5% of reading + 10V
Current Sourcing 30 to 40mA (above 500V, 180sec. max.)
0 to 29.99mA (above 500V, continuous)
0 to 10mA (below 500V, continuous)
Current Limit Range 0.10 to 40mA,
0.02mA/step
Current Accuracy 1.5% of reading + 50μA
DC Hipot
Voltage Range 0.100kV to 6.000kV
Voltage Setting 5 Volt Increments
Voltage Regulation 1% + 5V
(Line & Load)
Voltage Accuracy 1.5% of reading + 10V
Current Sourcing 0 to 10mA (above 500V, continuous)
0 to 2mA (below 500V, continuous)
Current Limit Range 0.10 to 10mA,
0.01mA/step
Current Accuracy 1.5% of reading + 50μA
Insulation Resistance
DC Output Voltage 50V / 100V 500V / 1000V
Resistance Accuracy 50V / 100V 1 to 50MΩ: 5% of reading
51 to 200MΩ: 10% of reading
201 to 1990MΩ: 20% of reading
Resistance Accuracy 500V / 1000V 1 to 500MΩ: 5% of reading
501 to 2000MΩ: 10% of reading
2001 to 9990MΩ: 20% of reading
Continuity Measurement
Test Current 100mA
Resistance Accuracy ±0.075Ω
Resistance Range 0.001Ω to 1.500Ω