Associated Research 6554SA Quadchek Electrical Safety Analyzer

Associated Research 6554SA Quadchek 4-in-1 Automated Electrical Safety Analyzer
Enlarge Image
Request Quote Print Page
Manual

The QUADCHEK Electrical Safety Compliance Analyzer is another step in the evolution of electrical safety testing instrumentation. The QUADCHEK provides in a single instrument for the three most common high voltage tests (AC Hipot, DC Hipot, Insulation Resistance) together with a high current Ground Bond Tester. The 4-in-1 capabilities of QUADCHEK make it versatile enough to test to UL,CSA,VDE,IEC,TUV,and other agency requirements.The built-in IEEE interface makes QUADCHEK an ideal choice for automation applications.
 

Functional Specifications:

DIELECTRIC WITHSTAND TEST MODE
Input 115 VAC (±15%),47-63 Hz,Single Phase
230 VAC (±15%),47-63 Hz,Single Phase
User Selectable
Output Rating 5kV @ 40mA,(AC &DC)
Output Adjustment 0 - 5kV @ 40 mA (AC &DC), 10 volt/step
0 - 40.0 mA (AC & DC), 0.1 mA/step
High Trip Range 0 .1 -40 .0 mA, (AC &DC)
Accuracy ± ((2%of setting +0 .02 mA)
Low Trip Range  0 .0 - 39.9 mA, (AC &DC)
Accuracy ± ((2%of setting +0 .02 mA)
FFailure Detector Audible &Visual 
Voltage Display  0 .00 to 5 .00 kV Full Scale, LED Display
Accuracy -
Reading ± ((2%of reading +1 count)
Setting ± ((2%of setting +5 volts)
Current Display 0.05 to 9 .99 mA,10.0 to 40.0 mA, LED Display
Accuracy -
Reading ± ((2%of reading +0.02 mA)
DC Output Ripple ≤4%RMS (5 kVDC,40 mA)
AC Output Wave Form Sine wave, Distortion: ≤1%
AC Output Frequency 50 or 60 Hz ± 100 PPM, User Selectable
Output Regulation 1%of setting from no load to full load
Dwell Timer 1 -999 seconds in 1 second increments or continuous, accuracy ± 0 ..1 seconds.
Ramp Timer 0 - 99 seconds in 1 second increments accuracy ± 0.1 seconds.

 

INSULATION RESISTANCE TEST MODE
Output Voltage Range 200 -1000 Volts, 1 volt/step
Voltage Metering 3 Digits,0.00 -1 .00 kV
Accuracy ± ((2%of reading + 1 count)
Resistance Metering 3 Digits, 9.99 GΩ Full Scale
Accuracy ± ((2%of reading + 1 count)
Limit Ranges HI - Limit range -0 -9999 MΩ
LO - Limit range -0 -9999 MΩ
Dwell Timer 1 - 999 seconds in 1 second increments
Accuracy ± 0.1 seconds

 

GROUND BOND TEST MODE
Output Voltage Adjustable (5.0 to 12.0 volts AC, 0.1 volt/step)
Setting Accuracy - ± (2%of setting +0 .1 volt)
Output Frequency 50 or 60 Hz, User Selectable
Output Current Adjustable (10.0 to 30.0 amps,0 .1 amp/step)
Current Display 3 digits,30.0 amps full scale, LED Display
Accuracy ± (2%of reading + 1 count)
Trip Range 1 to 600 mΩ
Dwell Timer

1 -999 seconds in I second increments or
continuous. Accuracy ± 0.1 seconds.