Pumps |
Dry Scroll |
|
Fore Pump Type |
IDP-3 |
|
Free Air Displacement,
l/min, 60/50 Hz |
60/50 |
Minimum Detectable Leak, atm-cc/sec |
5 x 10-12 |
Internal Pump Version |
IDP-3 |
Turbo Molecular Pump |
Navigator™ Turbo molecular pump specifically designed for leak detection |
Leak Units |
atm-cc/sec, mbar-l/sec, torr-l/sec, Pa-m3/sec, std. ft3/year |
Helium Response Time |
0.5 second, within 66% of peak reading within the given time. Per AVS 2.1. |
Test Port |
NW-25; A 11/8" (28 mm) ID Compression Port is available as an accessory. |
Recommended Operating Ambient Conditions |
Installation |
Indoor use only. Altitude up to 2000 m (6,400’), Material Group III, POLLUTION DEGREE 2 according to EN61010-1 Standard, Class 1 |
Operating Temperature |
System: +12° C to +45° C (wet and dry pumps); Fore pump limited on the low ends and turbo pump limited on the high ends. |
Humidity |
Maximum relative humidity (HR) 90% |
Heat Load |
1000 W (+/- 5%) |
Air Pressure |
563 Torr (75 kPa) to 795 Torr (106 kPa) |
Maximum system vibration |
3 G’s (Turbo Limit) |