Frequency range |
10 Hz - 3 kHz |
Sensor type |
Isotropic, RMS Combined measurement of electric and magnetic field |
Type of frequency response |
1) Weighted (Results displayed in % of the selected standard) 2) Flat response (Results in V/m, µT, etc. ) |
Exposure limits (probe in weighted mode) |
Public and occupational ICNIRP 2010 Customizable to other standards |
Measurement range |
Weighted mode (ICNIRP 2010) → E-field: 0.025 % - 200 % of limit (RMS value)
H-field: 0.025 % - 200 % of limit (RMS value)
Field Strength Mode → E field: 2.5 V/m - 20000 V/m (RMS)
H field: 0.05µT - 2000 µT (RMS) |
Dynamic range |
92 dB |
Sensitivity |
Weighted (E,H) 0.025 %
Flat response E field 2.5 V/m
Flat response H field 0.05 µT |
Frequency response |
± 20 % (typ.) of standard
(25 Hz - 1 kHz)
± 25 % (max.) |
Linearity |
± 1 % (typ.) (1 % - 100 % of standard)
± 2 % (max.) |
Isotropic response |
± 5 % (typ.) |
Calibration |
ISO 17025 Accredited Calibration (ILAC) |
Calibration period |
24 months (recommended) |
Operating temperature |
- 15 ºC a 50 ºC |
Dimensions |
270 mm x 115 mm Ø |
Field sensor area |
100 cm2 |
Weight |
210 g |