The Haefely PEFT 8010 instrument contains all the features expected from a top quality EFT generator. Unbeaten performance paired with a high end assembly guarantee a cost effective, long-lasting investment and valuable test results.

PEFT 8010 can either be operated by the front panel keys in a stand alone manner or be controlled by PC via RS232 or IEEE 488 as part of a complete EMC test system.


Features:

  • EFT impulse generation according to newest standards and latest technology.
  • ESD control and report functions for attaching an Electrostatic Discharge gun.
  • Automatic test operation for shortest measuring time and minimal setup effort.
  • Compact and reliable construction for factory, laboratory or field use.
  • Remote control is provided by optional software (WinFEAT&R) designed specially for EMC testing.
  • ISO 9001 : 2000 certified manufactured

 

Application Areas:

  • Approval Testing
  • Product Development
  • Post Design Diagnostics
  • Industrial
  • Medical
  • Telecommunication
  • Household

 

Specifications:

Wave Shape into 50 ohm load
Rise time 5 ns ± 30 %
Impulse duration 50 ns ± 30 %
Wave Shape into 1000 ohm load
Rise time 5 ns ± 30 %
Impulse duration 50 ns – 15 / + 100 ns
Burst Characteristic
Voltage at HV-output:
Voltage at CDN output:
1.0 … 7.3 kV
1.0 … 6.6 kV
Spike frequency 1 Hz … 110 kHz
Burst duration 0.01 … 999 ms
Burst period 0.1 Hz … 400 Hz
2.5 ms … 10 s
1 period … 500 periods
Impulses per second max. 250 for Unom > 4 kV
max. 500 for Unom ≤ 4 kV
Impulses per burst max. 75 for Unom > 4 kV
max. 150 for Unom ≤ 4 kV
Spike distribution normal, continuous, random, real
Polarity positive, negative
Test time 10 s … 8 h per path
Single Phase Coupling/Decoupling Network
Maximum AC voltage 264 V @ 16 … 440 Hz
Maximum AC current 16 A @ 50 … 60 Hz, 10 A @ 400 Hz
Maximum DC voltage 125 V
Coupling Modes L-GND, N-GND, PE-GND, LN-GND, LPE-GND, NPEGND, LNPE-GND
Residual voltage at Test Supply Input ≤ 10% of applied test voltage
Synchronisation
Frequency 16 2/3 Hz, 40 Hz, 50 Hz, 60 Hz, 400 Hz, auto
Impulse trigger automatic, manual, external
System Configuration
EUT fail input BNC, logic low to trigger
P90 interface for a 3-phase CDN extension
PESD interface for connecting an ESD generator
Printer interface Centronix
Computer interface RS-232, IEEE-488 (optional)
Power Supply
Voltage 85 … 264 VAC, 150 VA
Frequency 50 / 60 Hz
Weight and Dimension
Weight 14 kg
W x H x D 450mm x 130mm x 570mm