Range/Resolution Table

True RMS Voltage True RMS Current
.2000A 2.000A 20.00A
150.00V 30.00W 300.0W 3000W
300.0V 60.00W 600.0W 6000W
600.0V 120.00W 1200.0W 12000W
  True WATTS


Specifications

AC/DC Voltage & AC/DC Current (True RMS)
Range & Accuracy DC & 40 Hz to 5 kHz, ± 0.1% of reading ± 6 digits
5 kHz to 15 kHz, ± 0.5% of reading ± 6 digits
15 kHz to 20 kHz, ± 0.75% of reading ± 6 digits to 50 kHz with typical 1% error per 10 kHz
Crest Factor Response 50:1 for minimum RMS input, linearly decreasing to 2.5:1 for full scale RMS
Minimum Input 5% of range
Maximum Input 600V DC or RMS AC, 1500V peak
Maximum Common Mode 1500V peak, neutral to earth
Peak Indicator Illuminates at 2.5 times full scale
Maximum Input 35A peak, 20A DC or RMS; 100A DC or RMS for 16ms without damage
Overrange 150% of full scale for DC, up to maximum input
WATTS (True Power-EI COS Ø)
Range & Accuracy DC & 40 Hz to 5 kHz, ± 0.25% of reading ± 6 digits
5 kHz to 10 kHz, ± 0.5% of reading ± 0.5% of range
10 kHz to 20 kHz, ± 1% of reading ± 1% of range (2 A range only)
20 kHz to 50 kHz, additional ± 1% per 10 kHz above 20 kHz
Power Factor Response Unity to zero leading or lagging
Power Factor Resolution (.001 pf) zero to 1.000 pf (1000 counts)
Displays Dual 4½ digit
Load Connection 4 Terminal heavy duty binding posts
Size 9.25" L x 5" W x 8.5" H (23.5 cm x 21.6 cm x 6.4 cm)
Weight 5 lbs (2.3 kg)

Options:
  • 20mA - Low Current Version
  • I-150 - Clamp-on 150A CT
  • I-1000 - Clamp-on 1000A CT