Teseq KEMZ 801AS 10 kHz - 1 GHz EM Clamp for IEC/EC 61000-4-6

Rent Teseq KEMZ 801AS 10 kHz - 1 GHz EM Clamp for IEC/EC 61000-4-6
Enlarge Image
Request Quote Print Page
Datasheet
IEC 61000-4-6 (EN 61000-4-6) defines three basic types of transducer used for the injection of test signals into the equipment under test (EUT). The basic requirement is to inject a known level of RF signal onto the cable of the EUT at each test frequency and to determine whether the EUT continues to function correctly. In order to do this, it is necessary to decouple the auxiliary equipment (AE) from the test signal. If this is not achieved it is difficult to know whether any fault is due to a failure of the EUT or the AE. Whilst the coupling/decoupling network (CDN) is one of the best ways to achieve this test and indeed it is mandatory to use them under some circumstances, it is not always possible to use a CDN.

The KEMZ 801AS is supplied with calibration kit CAL 801A which allows test level setting and correction factor measurement.

Features:

  • As specified in IEC/EN 61000-4-6
  • Very efficient coupling
  • Can be used on almost any cable
  • Ruggedly designed

Specifications:

Parameter Value
Frequency range

10 kHz to 1000 MHz

Nominal impedance

50 Ω

Connector

N-type female

Maximum input level

0.01 to 100 MHz: 100 W for 15 min
100 to 230 MHz: 100 W for 3 min
230 to 1000 MHz: 50 W for 3 min

Maximum correction factor

0.15 to 400 MHz: 4 dB (50 Ω system), 13.5 dB (150 Ω system)